FiO2 sensor (with cable included)

006347

Accessory Type

Sensor

Product Tags

Vivo 50, Vivo 55, Vivo 60, Vivo 65

Quantity

1 Unit

Region

All

Description

Vivo 55/65 FiO2 Cell Technical Specification

Technical Specification
Measurement Range 0 to 100 Vol.%
Expected Operating Life – 1,000,000 Vol.% h
Sensor Lifetime < 6 years@ ambient air, depending on application
Initial Output Signal 13.0 to 16.5 mV@ dry ambient air
Response Time < 12 s
Drift < 1% Vol. O2/month@ air, averaged across 12 months
Operating Temperature 10 to 40 °C
Pressure Range 700 to 1250 hPa
Linearity Error W 3@ 100 % 02, applied for 5 min
Zero Offset Voltage 0,3 Vol% O2@ 100 % N2, applied for 5 min
Influence of Humidity – 0.03 % rel. O2 reading per% RH
Recommended Load Resistor: 10 kOhm
Temperature Compensation: NTC
Interferences: according to DIN EN ISO 80601-2-55
Weight 26g
Material in Contact with Media PA, PPS, PTFE, stainless steel

 

Storage Conditions
Temperature Range Recommended: 5 to 30 °C Maximum:- 20 to 50 °C ( lOh)
Humidity: up to 100 % RH, non-condensing
Shelf Life < 6 months recommended